Newheek specializes in producing X optical machine image intensifier, high voltage power supply,image intensifier TV system and so on.

HomeBlog ›Is blurring of X-ray machine image a damage to image intensifier TV System fluoroscopy ?

Is blurring of X-ray machine image a damage to image intensifier TV System fluoroscopy ?

Is blurring of X-ray machine image a damage to image intensifier TV System fluoroscopy
Some customers want us to help him repair the image intensifier TV System fluoroscopy. At present, the damage phenomenon is image blurring. In view of image blurring, we draw the following conclusions: the high voltage power supply of the image intensifier TV System fluoroscopy is damaged, or the output screen of the image intensifier is discharged, or the inner gallbladder of the image intensifier TV System fluoroscopy is leaked. In view of these situations, we suggest that the customer send his image intensifier to us for detection. 。
When we first examined the image intensifier TV System fluoroscopy, we found that the voltage of the high voltage power supply was unstable, which was the direct image intensifier. In this case, we recommend that customers change the high voltage power supply.
We have Toshiba, Thales and other image intensifier TV System fluoroscopy repair replacement products.
Is blurring of X-ray machine image a damage to image intensifier TV System fluoroscopy

Author:Abby

Contact us

Tel: (+86) 18953679166

Whatsapp: +86 18953679166

Email: service@newheek.com

Company: Weifang Newheek Electronic Technology Co., Ltd.

ADD: E Building of Future Star Scientific Innovation Industrial Zone of No.957 Wolong East Street, Yulong Community, Xincheng Sub-District Office, Weifang Hi-tech Zone, Shandong Province, China


(+86) 18953679166
service@newheek.com