Some customers want us to help him repair the image intensifier TV System fluoroscopy. At present, the damage phenomenon is image blurring. In view of image blurring, we draw the following conclusions: the high voltage power supply of the image intensifier TV System fluoroscopy is damaged, or the output screen of the image intensifier is discharged, or the inner gallbladder of the image intensifier TV System fluoroscopy is leaked. In view of these situations, we suggest that the customer send his image intensifier to us for detection. 。
When we first examined the image intensifier TV System fluoroscopy, we found that the voltage of the high voltage power supply was unstable, which was the direct image intensifier. In this case, we recommend that customers change the high voltage power supply.
We have Toshiba, Thales and other image intensifier TV System fluoroscopy repair replacement products.