Mobile image intensifier distortion fault and processing

Mobile image intensifier failure usually occurs in two cases: one is no image at all;The other is that geometric shapes can never be perfectly proportioned to the shapes of the corresponding objects. This phenomenon is often called image distortion. The main reason is that mobile image intensifier itself has two types of distortion.The other is asymmetry, often called S distortion.
Mobile image intensifier distortion fault is a common fault. When the ground line of video signal is disconnected, there will be a fault like integral distortion and maintenance of the distortion phenomenon.
Usually, the method of first testing power supply is normal, then to determine whether the video signal line ground wire is good, the final test mobile image intensifier high voltage driving circuit of each test point voltage is normal, if it is a mobile image intensifier vacuum part distortion caused by aging, is unable to repair, so the distortion of fault factors need to balance and make the correct maintenance plan.
In this regard, Newheek mobile image intensifier can provide quality detection and maintenance services. Can meet the needs of different customers.