Customer in Shanghai inquires about the image intensifier
Recently, a customer in Shanghai asked about the problem of the image intensifier, saying that the image intensifier could not produce images. Let’s talk about the phenomenon of poor vacuum in the image intensifier.
①Performance phenomenon: The main manifestation of the poor vacuum of the inner tank is: when the intensifier is energized, without radiation, the output screen of the intensifier will show a semi-crescent light. The larger the semi-crescent light area, the vacuum of the inner tank The bad state is more serious.
②Solution:
First: If the vacuum defect is not very serious, you can carry out a period of aging test (provide power for the intensifier for a long time for 1-3 days) to observe whether this phenomenon still exists.
Second: If the vacuum phenomenon is more serious, only replace the inner tank or purchase new products.
If you have questions about other image intensifiers, you can also contact us.