E5804 image intensifier is equipped with NK5761HD-P1P5 high voltage power supply

Henan ** Health Center purchased newheek NK5761HD-P1P5 X-ray high-voltage power supply with E5804 image intensifier. The image is not very clear because the engineer does not know how to install and debug. In addition, the X-ray image intensifier is installed in the basement. Discharge phenomenon. Similar to this situation, it is recommended that customers mail the entire X-ray image intensifier directly, our company will do a comprehensive inspection for free, and install and debug the X-ray high-voltage power supply.

The precautions for the replacement and installation of the X-ray high-voltage power supply are as follows:

1. The precautions mentioned below must be carried out on the premise that the inner liner of the X-ray image intensifier can be used normally.

2. Before installation, please clean the high-voltage line connected to the X-ray high-voltage power supply on the X-ray image intensifier with alcohol and dry it to prevent the high-voltage line from being dirty and causing discharge after installation.

3. When installing, connect the cleaned high-voltage line to the corresponding high-voltage output port of the X-ray high-voltage power supply according to the line number. Remember that the high-voltage wire must be inserted to the end when connecting, and the high-voltage wire must be fastened. In order to avoid the loosening of the high-voltage line and sparking.

4. If the high-voltage line is also cleaned; it is also fastened, but there is still a slight discharge at the high-voltage end. The high-voltage line can be pulled out when the power is off, and the discharge problem will be solved after applying some high-voltage silicone grease on the line and reconnecting it.

5. In the process of adjusting the image, you can debug according to the method mentioned in the manual.

If the user cannot successfully install the high-voltage power supply according to the X-ray high-voltage power supply manual and installation precautions, it is recommended to mail the X-ray image intensifier to our company. We will install and debug it for free.

If you still don’t understand, please get in touch with us by E-mail!