Not long ago, the X-ray high-voltage power supply of an image intensifier in a hospital ignited when it was turned on. The customer found our company on the Alibaba platform and asked if the situation could be repaired. The customer’s X-ray high-voltage power supply model is as follows:
In view of the customer’s situation, because we are not at the scene, we have not seen the malfunction. If the customer wants to repair it, it is recommended to mail it directly. Our company is responsible for detecting the cause of the failure free of charge, and judging whether it can be repaired based on the result. In addition to the X-ray high-voltage power supply, the image intensifier has a key component of the inner tank. If the X-ray high-voltage power supply is on fire for a long time, the inner tank may also be burned out. Therefore, if the power supply is damaged, verify it. Whether the inner tank is damaged. The customer said that the engineer had checked it, and the inner tank was okay, but the image flickered a little when the X-ray high-voltage power supply was on fire. The sparking point of the X-ray high-voltage power supply is mainly the red high-voltage end in the picture.
At present, our company has received the X-ray high-voltage power supply mailed by the customer. After the inspection by our engineers, the high-voltage end has been carbonized and cannot be repaired. The customer asked: Can’t you change the high-voltage end? Tell the customer because the internal burnout, even if the high-voltage end is changed, it will not help. The customer asked again: What caused the internal burn through? It was reported that because the X-ray high-voltage power supply had a slight discharge and was not treated in time, it was carbonized and burned through. In this case, if you continue to use it, the inner liner is also easy to burn out. The customer said it understood.
Dear users, if your equipment has this kind of problem, remember to repair it in time. The best way is to replace it with a new X-ray high-voltage power supply in time. Otherwise, the longer it is, the other parts will be damaged. If you have any needs, remember to contact us!