Image intensifier – 9 inches
1. Model: NK-23XZ
2. Purpose: The X-ray image intensifier (hereinafter referred to as the intensifier) converts X-ray image input into visible light image output
Electronic equipment designed for installation on X-ray TV devices, used for X-ray fluoroscopy and photography.
3. Structure: The X-ray image intensifier consists of an X-ray image intensifying tube, a tube container, and a dedicated high-voltage power supply housed within the tube container
and the composition of the low-voltage power supply driving the high-voltage power supply.
4. Product performance parameters
| Effective incident field size | 215 ± 3mm | 160 ± 2mm | 120 ± 2mm | unit |
| Cathode | 0 | 0 | 0 | V |
| E1 | 60 | 121 | 78 | V |
| E2 | 612 | 305 | 368 | V |
| E3 | 2228 | 4710 | 7290 | V |
| Anode | 30000 | 30000 | 30000 | V |
| Output Image Diameter | 25±0.5 | – | – | mm |
| Central Resolution | ≥52 | ≥58 | ≥68 | Lp/cm |
| Input Voltage | 24±1 | – | – | V (DC) |
5. Replacement for Thales (Thomson) TH9438QX, TH9438HX, TH59438 or Toshiba E5830HD, E5830SD
Image intensifiers from Siemens, OEC, etc
6″, 9″, 6″, and 4.5″ support three-view switching
Newheek NK-23XZ Image Intensifier 9″
1.Brief Introduction:
Newheek NK-23XZ X-Ray Image Intensifier is the electronic output equipment which converts the x-ray image into the visible light image. It is installed on the x-ray TV system, which is applicable to the x-ray fluoroscopy and radiograph.
2. Construction:
Newheek NK-23XZ X-Ray Image Intensifier is composed of x-ray image multiplier tube, tube container, high voltage power supply which is set inside of tube container and the low voltage power supply which drives the high voltage power supply.
3. Application:
The nominal entrance field of view of Newheek NK-23XZ image intensifier is 23cm (9inch), which is mainly application to C-arm, multi-functional stomach and intestine radiograph and fluoroscopy x-ray machine, digital RF x-ray machine, lithotrity and industrial detection X-ray machines etc.
4. Main Technological Parameters:
(1) Photoelectric Performance:
Input screen size: 230mm
Effective entrance field of view size: 215mm
Output image diameter: 20mm
Limiting resolution: 52Lp/cm
(2) Low Voltage Power Supply Performance Parameters:
Input voltage: 86V ~ 265V
Input power supply frequency: 50Hz/60Hz
Output voltage: 24V±0.5V
Output current (effective value): 1.5A


Tel: (+86) 18953679166