Customers have thought twice about claiming that medical scanner is very expensive after all, for fear of accidents during transportation and further damage, they decide to carry the image intensifier on site for inspection and maintenance.
Two customers came to our company, our company engineers to do the test, found the high voltage line at the ignition, decided to remove the whole high voltage line replacement, so as to solve the problem from the source.The customer agreed to repair, after the completion of re-packaging, the discharge phenomenon is significantly reduced, telling us that the current weather is relatively cold glue needs longer time to solidify, the effect will be better.After 30 hours of solidification, the ignition basically disappeared.
After the customer goes back to install, the glue is solidified. After installation and debugging, all images are normal.