Two most common problems and solutions for image intensifiers

Two most common problems and solutions for image intensifiers:

Fluoroscopy image intensifier system application Night vision image intensifier app
1. Leakage of the liner:
1 performance phenomenon: the main form of the image is blurred, the high voltage terminal voltage drops linearly.
2 solutions:
The image intensifier liner is a vacuum device. Once the liner is discharged, only the liner can be replaced or a new product can be purchased.
2. Poor vacuum in the bladder:
1 performance phenomenon: the main performance of the vacuum of the inner tank is: the booster is energized, in the case of no radiation, the output screen of the intensifier will appear half moon crescent, the larger the light area of the semilunar, the inner vacuum The worse the bad state.
2 solutions:
First: If the vacuum is not serious, you can carry out a aging test for a period of time (long-term power supply to the intensifier for 1-3 days) to see if this phenomenon still exists.
Second: If the vacuum is serious, you can only replace the liner or purchase new products.