Newheek specializes in producing X optical machine image intensifier, high voltage power supply,image intensifier TV system and so on.

HomeBlog ›How to measure the high-voltage power supply of the image intensifier

How to measure the high-voltage power supply of the image intensifier

A customer after-sales service department consults how to measure the high-voltage power supply of the image intensifier, and the response requires a DC multimeter to measure. One of the common failures of image intensifiers is a problem with the high-voltage power supply of the image intensifier.
How to measure the high-voltage power supply of the image intensifier, the specific operation method is as follows:
1. First use a DC multimeter to check the low-voltage end of the image intensifier: if the voltage is normal, the high-voltage board is damaged; if there is no voltage, the low-voltage board is broken.
2. Re-measure the high-pressure end of the image intensifier: if the high-pressure end is normal, the inner tank is broken; if the high-pressure end is abnormal, the high-pressure plate is broken.
How to measure the high-voltage power supply of the image intensifier is a very professional operation. Unless it is an engineer in the after-sales maintenance department, we generally recommend that customers send the intensifier to us for professional inspection.

 high-voltage power supply

Author:Lillian


(+86) 18953613955
service@newheek.com